API Command Reference
Appendix A
A-22 NetScan Users Manual
E?
- Error Status Query
TYPE
System
EXECUTION
Immediate
SYNTAX
E?
Returns present error condition of the unit with one of the following error
codes:
E000
- No error has occurred.
E001
- Invalid device dependent command (IDDC).
E002
- Invalid device dependent command option (IDDCO).
E004
- Channel configuration error.
E008
- Calibration error.
E016
- Trigger overrun.
E032
- Open T/C or Range Error.
E128
- Command conflict error.
DESCRIPTION
When an error has occurred, the ERROR indicator light on the front panel of the unit turns on. Error Status
Query (
E?
) is used to determine the present error condition. After execution of the Error Status Query command,
the unit responds with one of the following error codes:
E000
: No error has occurred.
E001
: Invalid device dependent command (IDDC) due to a command syntax error.
E002
: Invalid device dependent command option (IDDCO). A command parameter was out of range or
missing.
E004
: A Channel Configuration Error indicates that a channel has been inappropriately configured, either
because the channel is not present or because the specified channel type is not compatible with the card type
installed.
E008
: A Calibration Error occurs when the card calibration has failed or has been inappropriately calibrated.
E016
: A Trigger Overrun Error occurs when more than one trigger event occurs for a given trigger state or when
a non-expectant trigger event occurs.
E032
: An Open T/C Error or Range Error, occurs when the hardware circuitry detects an open thermocouple, or
when the software detects the A/D input has reached or exceeded its linearization.
E128
: A Command Conflict Error indicates the issuance of a command that is in conflict with other commands
that have been issued or the current configuration. This error usually occurs when a command is issued that
cannot be performed because of the current state of the unit. For instance, if the Set Scan Interval (
I
) command
is issued during an acquisition (
tstart 0
).
After execution of the Error Status Query, most error conditions are cleared. Error conditions may also be
cleared by issuing a Clear Event Status (
U0
) command to the unit. The ERROR light will remain on until an
E?
or
U0
command is executed to clear the error condition.
EXAMPLES
PRINT#1,"E?X"
Request error status
INPUT#1,A$
PRINT A$
Display shows
E000
, no errors occurred
PRINT#1,"K3 X"
Send illegal command option to the unit
(Note: ERROR indicator light turns on)
PRINT#1,"E?X"
Request error status
(Note: ERROR indicator light turns off)
INPUT#1,A$
PRINT A$
Displays
E002
(Invalid Device Dependent Command Option)